App DB | Sorbent Technologies, Inc.
Home
sorbtech.com
Back
Contact an Account Manager
Determination of aminoglutethimide and N-acetylaminoglutethimide in human plasma by HPLC
Summary
Type:
HPLC
Application #:
104510
Catalog#:
720141.46 - MN HPLC Column, Nucleosil C18, 100A, 5um, 4.6x100mm
Detection:
UV, 235nm
Conditions:
eluent: acetonitrile - methanol - water (5:20:75, v/v)
flow rate 1ml/min
temperature 30°C
Authors:
Menge G
Source:
J Chromatogr 310 (1984) 431 - 437
Details
Legend: Fig. 2. Chromatograms of A) blank plasma sample containing internal standard, B) spiked plasma sample containing 0.4µg/g each of aminoglutethimide (AG) and N-acetylaminoglutethimide (N-AcAG), C) pla